Specification
Filtration: Rechargeable Lithium Ion battery; AC power using a protectively grounded outlet
Approximate Dimensions: 23.5 cm (H) x 29 cm (W) x 20.5 cm (D) 9.25 in (H) x 11.4 in (W) x 8 in (D)
Approximate Weight (without battery): 5.66 kg; 12.5 lbs
Standard Operator Position: Within one meter (3 feet) of the device
Power: Rechargeable Lithium Ion battery; AC power using a protectively grounded outlet
Size: Approximately 7 in (17.8 cm) diagonal viewing area
Type: Color TFT LCD
Resolution: 800 x 480 pixels (VGA) with 32 brightness levels per color
Sweep Speed: 25 mm/s ± 10% nominal (stationary trace; sweeping erase bar) for ECG and SpO2; capnogram wave is 6.25 mm/s ± 10%
Wave Viewing Time: 6.5 sec ± 10%
Waveform: Biphasic Truncated Exponential; waveform parameters adjusted as a function of patient impedance
Shock Delivery: Via multifunction electrode pads or paddles
Shock Series: Configurable energy escalation in a series Leads Off Sensing and PCI
Charge time:
• Less than 5 seconds to the recommended adult energy level (150 Joules) with a new, fully-charged battery installed
• Less than 6 seconds to the selected energy level (up to 200 Joules) with a new, fully charged battery installed
• Less than 15 seconds to the selected energy level while connected to AC power only
• The device powers on in manual defibrillation mode ready to deliver shock in less than 15 seconds with a new, fully-charged battery
• The device powers on in AED mode ready to deliver shock in less than 24 seconds with a new, fully-charged battery even after 15 discharges of maximum energy
Sensing for Pads/Paddles: Apply 500 nA rms (571 Hz); 200 uA rms (32 KHz)
Charge times:
• Less than 5 seconds to the recommended adult energy level (150 Joules) with a new, fully-charged battery installed
• Less than 6 seconds to the selected energy level (up to 200 Joules) with a new, fully charged battery installed
• Less than 15 seconds to the selected energy level while connected to AC power only
• The device powers on in manual defibrillation mode ready to deliver shock in less than
• 15 seconds with a new, fully-charged battery
• The device powers on in AED mode ready to deliver shock in less than
• 24 seconds with a new, fully-charged battery even after 15 discharges of maximum energy
Patient Impedance Range: Minimum: 25 ohm (external defibrillation); 15 ohm (internal defibrillation) - Maximum: 250 ohm. Actual functional range may exceed these values