Specification
Item Ref. WD7PAT487
Temperature module: Genuis™ temperature modul
SpO₂ module: seca SpO₂ module
Power supply: Built-in power adapter, internal rechargeable lithium-ion battery
Net weight: 0.4 sts, 2.7 kg
Interfaces: Ethernet, USB 2.0, Wi-Fi
Dimentions: (WxHxD) 252 mm, 262 mm, 278 mm
Display: 7" Touchscreen-Display
Blood pressure measurement method: Oscillometric Measurement
Measurement accuracy at ambient temperature: 16 °C – 33 °C (60.8 °F – 91.4 °F) Target temperature: 36.7 °C 38.9 °C (98.1 °F 102 °F) ± 0.1 °C ( ± 0.2 °F)
Measurement accuracy at target temperature: 25 °C (77 °F): Target temperature: 33 °C 42 °C (91.4 °F 107.6 °F) ± 0.2 °C ( ± 0.4 °F)
Blood pressure measurement certainty: Excess pressure limit 300 mmHg;
automatic pressure release at > 300 mmHg
Blood pressure measurement time: Normal: 15 - 20 sec / max. 90 sec (adults)
Blood pressure measurement range – deflation: pSYS: 25 - 280 mmHg pDIA: 10 - 220 mmHg pMAP: 15 - 260 mmHg
Blood pressure measurement range – inflation: pSYS: 77 - 200 mmHg pDIA: 45 - 190 mmHg pMAP: 56 - 193 mmHg
Blood pressure measurement accuracy – inflation (clinically validated according to ISO 81060): Mean deviation systole 0.94 mmHg; Standard deviation systole 3.84 mmHg; Mean deviation diastole 0.57 mmHg; Standard deviation diastole 3.17 mmHg
Blood pressure measurement accuracy – deflation (clinically validated according to ISO 81060): Mean deviation systole 0.39 mmHg; Standard deviation systole 2.57 mmHg; Mean deviation diastole 0.43 mmHg; Standard deviation diastole 1.73 mmHg
Blood pressure pulse rate range – deflation: 30 - 240 bpm
Blood pressure pulse rate range – inflation: 45 - 200 bpm
Blood pressure pulse rate accuracy: max. +/- 3 bpm
SpO₂ measurement method: Pulse oximetry
SpO₂ measurement range: 0 - 100 %
SpO₂ measurement accuracy: SpO₂ - resting (60 % - 100 %) ± 2 Arms; SpO₂ - in motion (70 % - 100 %) ± 3 Arms
SpO₂ pulse rate measurement range: Pulse rate PR (Standard) 30 bpm to 240 bpm; Pulse rate PR (Enhanced) 20 bpm to 300 bpm
SpO₂ pulse rate measurement accuracy: ± 2 bpm
Temperature measuring method: in Ear
Temperature measuring range: 33 - 42 °C / 91.4 - 107.6 °F
Temperature measurement response time: < 2 sec.